AP034 力科有源差分探頭
- 公司名稱 蘇州眾業(yè)電子有限公司
- 品牌
- 型號 AP034
- 產地 美國
- 廠商性質 代理商
- 更新時間 2017/6/25 1:50:22
- 訪問次數 1384
聯系我們時請說明是化工儀器網上看到的信息,謝謝!
力科有源差分探頭英文簡介:
Fully Integrated: With the ProBus™ interface, the AP033 becomes an integral part of the oscilloscope. The probe sensitivity and offset can be controlled from the oscilloscope front panel, the probe front panel, or by using remote control commands (GPIB or RS-232). Sensitivity, offset, input capacitance, and common mode voltage range are displayed on the oscilloscope screen. When used with a LeCroy digital oscilloscope, no external power supply is required.
Wide Dynamic Range: The AP033 provides a range of sensitivities from x10 gain to ÷10 attenuation (even ÷100 with plug-on attenuator) for diverse signals. The sensitivity can be adjusted continuously from 200 µV/div to 1 V/div, when used with a LeCroy oscilloscope.
At high sensitivities, the x10 GAIN and the ÷1 ATTENUATION are selected. This produces 4 nV/√Hz input noise, ±5 V of common mode voltage range (CMVR), and ±400 mV of DC differential offset. At lower sensitivities, the x1 GAIN and ÷10 ATTENUATION are selected, resulting in ±50 V CMVR and ±4 V of differential offset.
The input capacitance is just 2 pF/side (1 pF differential) whenever the ÷10 ATTENUATION is selected. Deflection factors of 2 mV/div to 1 V/div can be obtained with only 2 pF/side input capacitance and ±50 V CMVR.
DC CMRR is greater than 10,000:1 (80 dB) in both ATTENUATION settings.
OFFSET up/down buttons allow OFFSET control from the probe amplifier body. Momentarily holding both buttons will zero the OFFSET.
A supplied AC coupling capacitor head (0.1 µF) allows operation with large common mode or differential DC inputs.
The input attenuator automatically changes from ÷1 to ÷10 when too much common mode signal is sensed. This and a fast input protection circuit also serve to protect the input from potentially damaging signals. When the probe is removed from the oscilloscope or the power is turned off, the ÷10 ATTENUATION is automatically invoked, leaving the inputs protected when not in use.
Gain and Attenuation Selections
The AP033 consists of a probe head with ÷1 or ÷10 attenuation and an amplifier body which attaches to the scope front panel. The amplifier gain can be either x1 and x10. Both the probe head attenuator and amplifier gain are controlled through the ProBus interface. Normally the GAIN and ATTENUATION selections are accomplished by the oscilloscope as shown in the table below.
There are two ways of obtaining x1 gain from the AP033: x1 GAIN with ÷1 ATTENUATION, or x10 GAIN with ÷10 ATTENUATION. The oscilloscope selects the latter of these two options (as shown below) to provide minimum input capacitance and maximum common-mode dynamic range while still leaving low noise operation available at the highest sensitivity.
By using the probe amplifier body control to override the automatic selection, the x1 GAIN with ÷1 ATTENUATION option is readily available for 10 mV, 20 mV and 50 mV/div. CMVR is ±5 V, input capacitance is 5 pF, and noise is 9 nV/√Hz. Similarly, x10 GAIN and ÷10 ATTENUATION can be selected at 2 mV/div and 5 mV/div for 2 pF input C and ±50 V CMVR. The oscilloscope displays the correct deflection factor in all cases.
Using the supplied accessory plug-on attenuator head, gain selections from 100 mV/div to 10 V/div are available.
Input Characteristics
The input capacitance is modeled by 2 pF to ground from each input plus 0.20 pF from input to input when the ÷10 ATTENUATION is selected. This is equivalent to 1.20 pF differential input capacitance. With ÷1 ATTENUATION, the differential input capacitance is 2.7 pF.
Input resistance is 1 Mohm at all GAIN and ATTENUATOR settings, including the accessory plug-on attenuator head.
An AC coupling plug-on accessory features a 0.1 µF coupling capacitor for each input. The low-frequency CMRR is reduced.
Autobalance
Holding both offset buttons down for two seconds with the input connections removed invokes Autobalance. This provides the highest accuracy on all ranges by removing residual DC offset from the probe.
Low Noise
At full gain, the noise is less than 4 nV/√Hz, the lowest of any available differential probe.
Common Mode Sensing and Input Protection
It is not uncommon for a differential amplifier to be connected to a signal whose common- mode voltage exceeds the amplifier''s common-mode range. Faulty readings, including total loss of signal, can result. The AP033 senses when signals exceed ±5.5 V and switches its input attenuator into the signal path. The oscilloscope shows the new deflection factor, and the probe continues to work.
The same mechanism protects the probe from damage due to large input signals. The probe amplifier is protected from large, fast-rising signals until the input attenuator can provide permanent protection. This takes about 1 ms. With the input attenuator in place, the input can withstand 200 V DC indefiniy.
力科有源差分探頭 圖片展示:
產品實圖 |